Blood Pressure Estimation Using Oscillometric Pulse Morphology

Blood Pressure Estimation Using Oscillometric Pulse Morphology

Show full item record

Title: Blood Pressure Estimation Using Oscillometric Pulse Morphology
Author: Mafi, Majid
Abstract: This thesis work presents the analysis of Oscillometric blood pressure pulse waveform under different pressure points (Systolic, Mean Arterial, and Diastolic Pressures). Pulse waveforms' characteristics were determined from the waveforms at three different pressures and are compared for subjects at three different age groups. Estimation of blood pressure using a morphology based approach was done by using the change of pulse waveform characteristics at different pressure points. Pulse waveforms' characteristics that were obtained from pulse waveforms are utilized to estimate SBP, MAP, and DBP. The estimates obtained with pulse morphology based technique are compared with a BP measurement device and Maximum Amplitude Algorithm. Maximum slope of the pulse was also used for blood pressure estimation. The effect of movement and breathing on proposed method and MAA were compared and it was observed that breathing artifacts affect less the proposed method.
Date: 2012
URI: http://hdl.handle.net/10393/20630
Supervisor: Groza, Voicu
Bolic, Miodrag
Faculty: Génie / Engineering
Degree: MASc

Files in this item

Files Size Format View
Mafi_Majid_2012_thesis.pdf 2.950Mb application/pdf View/Open

This item appears in the following Collection(s)

Show full item record


Contact information

Morisset Hall (map)
65 University Private
Ottawa ON Canada
K1N 6N5

Tel. 613-562-5800 (4563)
Fax 613-562-5195

ruor@uottawa.ca